10 lead 12 channel simultaneous 10 second resting ECG
296 (w) x 305.5 (h) x 92.5 (d)mm Approximately 3.5kg
3, 6, 12 channel and 1 channel (60 sec)
5, 10, 20, auto (I~aVF: 10, V1~V6: 5) mm/mV
12.5, 25, 50 mm/s
AC (50/60Hz, - 20dB or better) Muscle (25~35Hz, -3dB or better) Base line drift (0.1Hz, -3dB or better) Low pass filter: off, 40Hz, 100Hz, 150 Hz
320 x 240 character LCD display (3 leads trends)
ID, date, sensitivity, speed, filter, form, rhythm lead
Touch screen (alphanumeric and symbol available), keyboard, rotary push-knob (pop-up menu)
ID, name, age, sex, height, weight
Heart rate, PR, QRS, QT/QTc, P-R-T axis
Thermal print head, thermal roll paper Report papers - Width: A4: 210mm or 8.5" - Length: A4: 300mm or 11" Effective writing width: 204mm Resolution - Vertical: 8 dot/mm - Horizontal: 16 dot/mm
Internal noise : 20uV (p-p )max Input circuit : Floating input Isolated and defibrillation protected Input impedance : ¡Ý10M¦¸ Input range : ¡Ý¡À5mV Common Mode Rejection : ¡Ý+100dB DC offset voltage : ¡Ý¡À300mV Time constant : 3.2sec Patient leakage current : <10uA Frequency response 0.05~150Hz
Disconnect lead detection
Power supply: AC or built-in battery (option) 100-240 vac, 50/60Hz, 1.0-0.5 A, 60W max
1 hour of normal use (approx. 100 ECG printouts)
PC connection with RS-232 interface and LAN
Class 1, Type BF
CE, CSA
Operating humidity: 30~85% Operating temperature: 10¡ã ~ 40¡ã C Atmospheric pressure: 70 ~ 106 KPa
Power Cord - 1ea. Patient Cable - 1 ea. Electrode - 10 ea. Chart paper - 1 ea. Operation manual - 1 ea.
Rechargeable battery Cart Hanger Bag PC s/w Limb electrode Chest electrode RS-232C null cable LAN cable (ethernet-RJ-45)